ShuttersΒΆ

2-BM has two x-ray shutters:

  • Front End shutter upstream of 2-BM-A

  • Mono shutter upstream of 2-BM-B

If you are not able to open the 2-BM-A or 2-BM-B shutters check the following:

  • Shutter permit is enabled

  • To open the B shutter: A and B stations are searched

  • To open the A shutter: A station is searched

If you still cannot open the shutters check that the Equipment Protection System (EPS) is clear of errors. The EPS most frequent trips are caused by:

  • Low He flow caused by low He pressure

  • Vacuum bump casued by filter A moving with A shutter open

  • Mirror high temperature caused by A slits too wide vertically or mirror to high in the beam