ShuttersΒΆ
2-BM has two x-ray shutters:
Front End shutter upstream of 2-BM-A
Mono shutter upstream of 2-BM-B
If you are not able to open the 2-BM-A or 2-BM-B shutters check the following:
Shutter permit is enabled
To open the B shutter: A and B stations are searched
To open the A shutter: A station is searched
If you still cannot open the shutters check that the Equipment Protection System (EPS) is clear of errors. The EPS most frequent trips are caused by:
Low He flow caused by low He pressure
Vacuum bump casued by filter A moving with A shutter open
Mirror high temperature caused by A slits too wide vertically or mirror to high in the beam