Oscilloscope setup¶
To operate 2-BM Digita Oscilloscope under epics/medm you neet to start a soft epics IOC with:
[user2bmb@arcturus]$ cd /net/s6dserv/xorApps/epics/synApps_5_8/ioc/dlm/iocBoot/iocLinux
[user2bmb@arcturus]$ run
then in a different xterm start the medm screen with:
[user2bmb@arcturus]$ cd /net/s6dserv/xorApps/epics/synApps_5_8/ioc/dlm
[user2bmb@arcturus]$ start_caQtDM
and select
DLM400 button to start the control screen for the scope then
- caputRec
click on (re)start recorder
click on Refresh Menu
click Select Macro to pick dlm or dummy
click DO to collect a plot
to save the plots use
- scanH (for Harware-assisted scans)
set directory/file name where to save the data (in mda format)
if caputRec is set with dlm macro it collect multiple plots
if caputRec is set with dummy macro it ….
To view the mda files collected by the scan record use:
[user2bmb@arcturus]$ /home/beams/MOONEY/bin/mdaExplorer
or:
[user2bmb@arcturus]$ /APSshare/bin/dview
To convert MDA files to ASCII:
[user2bmb@arcturus]$ /APSshare/bin/mda2ascii